Info
The Signal Modules provide the interface to the different
signal sources to be measured. A range of individual Signal
Modules can be used for signal conditioning of different sensors
or data sources. A maximum of 24 channels can be connected to
one Signal Module. Visualisation of the input signals is carried
out by means of a display and overload LED’s. The display
ensures the identificatin of the module within a decentralised
configuration by displaying the selected unique module number.
The Signal Module performs the signal conditioning and the
A/D conversion (24bit). Each channel is completely independent
and consists of sensor support (power), amplifier, A/D converter,
anti aliasing filter and optional high or low pass filter. In
view of accuracy, measurement speed, and noise characteristic
the Signal Modules represent “State-of-the-art” technology. The
digital and the analogue module parts and the power voltage
input part are galvanic isolated to remove the noise introduced
by potential differences. All modules have their own power
supplies as well as their own calibration units. The low power
consumption of a Signal Module and the enhanced cooling
technology avoids the integration of fans in the Signal module
which is a big advantage especially for acoustic measurements.
The sample rate can be set for each module individually. A
special 6 channel module feature is the overload detection. An
overload situation will be detected by means of a comparator
already before the A/D conversion. Also one channel can be used
as signal channel or a RPM channel.
One Signal Module can be connected directly to the USB interface
of a PC.
Specs
Number of Channels : 24
Sampling Frequency (Fs ) : up to 50 kHz, settable in steps of 10
Hz. Each module can work with an own sampling frequency,
which is a binary divider of system FS
Bandwidth : to max. 20 kHz (settable)
Sampling Ratio : 2.5; no mirroring effect
Dynamic Range (SFDR) : > 90 dB at 0dBV Range in DATaRec 3
Mode
Analogue to digital converter : 24 bit
Data resolution : 16 bit ( DATaRec 3 Mode ) or
32-bit*( DATaRec 4 Mode ), selectable
THD : <-90 dB at 0dBV Range
Phase Accuracy : <0.2°
AC Accuracy : ±0.1 % or ±1 mV
DC Accuracy : ±0.1 % or ±1 mV
Input impedance : 1 M Ohm
Signal Level : ±10mV peak to ±10V peak
Coupling : AC (-3dB at 0.75 Hz typical)
ICP power : 22V DC, 4 mA constant current
TEDS sensor identification : IEEE 1451.4 standard for all
24 channels
RPM channel : channel 24 is useable as RPM input
RPM sampling frequency : DATaRec 3 – Mode :16 x Fs ( max.
800 kHz )
Threshold level : 0 – 100% in 5 % steps of the selected
amplifier step Mode : Single Ended
Connector : 1 * 50 pin D-sub
Dimensions : 48.1x 184 x 124,5 mm (w x h x d) including
fixing system
Weight : 1100 g typical
Cooling : no fan
Power consumption : 15 watts typical
Power input range : 17 up to 28 voltages
Input DC Power : Low Power detection
Communication : built-in USB 2.0, RS-232 and high speed
Link interfaces Functions:
Integrated calibration unit : 1 ppm reference Calibration
Galvanic isolation : yes
Max. isolation voltages : power input (48 voltage)
digital / analogue part (48 voltage)
Input channel voltage safety : ± 35 V peak
Environmental specifications:
Vibration : 5g
Shock : 10g
Operating temp. : -30°C - +65°C
Storage temp. : -40°C - +90°C
Humidity : 0 - 95% relative, non-condensing
|